Installation

  1. Define the position of the joystick.
    When the arrow at the bottom of the joystick points to the front, a forward movement of the joystick will result in an according forward movement of the wheelchair.
    If needed, you can change this positioning in steps of 90° with the Configurator Software.
  2. Place the mounting ring around the joystick housing.
  3. Secure the Q2M half clamp to the mounting ring.
    Use an Allen wrench to mount the 20 mm bolt.
    Note: It is not advised, but you may mount the joystick without using the Q2M half clamp.
    CAUTION: Tighten the screws firmly, but not excessively. Excessive force may damage the unit.
  4. Slide the unit over a Q2M Rod, or any other 6 mm rod, to position the Micro Joystick on the wheelchair.

    A Q2M Rod D6 50 mm is included in your package, but you can use any other Q2M Rod (e.g. a C- or S-Rod from the Multi Swing).

  5. Place and secure the interface unit and all cabling on the wheelchair.
    Warning: Place the Interface unit with the connectors facing downwards. This is to prevent water from entering the Interface unit.
  6. Place a power on/off (pwr) and/or mode (in) switch, secure their cabling and insert their connections. You can use the connections on the Interface unit.
    CAUTION: Before inserting a connector, remove the protective cover. If the connections are not used, always put or keep the protective covers in.
    CAUTION: This is Class I Medical Device (MDR 2017/745). All accessories, including switches, must also comply with the MDR 2017/745 regulations (e.g. the mo-vis Twister, tested according to EN12184 standards).
  7. Connect the cabling to the wheelchair electronics.
    CAUTION: All wheelchair electronics must be switched off during installation.